LAUDA Integral T 4600
Process therm. 400 V; 3/N/PE 50 Hz
Process thermostat with integrated cooling system for dynamic temperature control in external circuits
2-line back-lit LCD display for parallel display of actual and set temperature with messages of operating states in clear text
User-friendly menu guidance with simplest 3-key operation
Fully electronic continuous controller with PID action for internal and external control
Low-level and adjustable over-temperature protection with acoustic alarm for use with flammable and non-flammable liquids
Extremely powerful pressure pump
RS 232 and RS 485 interfaces and analogue inputs and outputs
Adjustable bypass for pressure limiting
Additional pump for internal circulation
Programmer max. 150 segments, divided into 5 programs
Digital pump pressure display, available at the analogue and digital interfaces
Stand-by circuit
Remote fault indication through floating contact
Filler opening at the front, drain tap
Small stainless steel bath with high expansion volume
Condenser cooling Air
Automatic proportional cooling with compressor on/off control
Power cord with plug (IEC 60309, 5-pol, CEE, red, 16 A)
Technical data (according to DIN 12876)
Working temperature range -30 °C … 120 °C
Ambient temperature range 5 °C … 40 °C
Temperature control ± 0,20 °C
Heater power 6,0 kW
Cooling capacity at 20 °C 4,40 kW
Power consumption max. 8,5 kW
Rated current 15,0 A
Circuit breaker T16
Pressure pump
– Pressure max. 3,2 bar
– Flow max. 40 L/min
Pump connections G 3/4
Filling volume min./max. 6,0 / 18,0 L
Overall dimensions (W x D x H) 550 x 650 x 970 mm